![]() |
Our first test was to see if the DCR exhibited a non-linear response
when observing sources of varying strength. We took data at the
positions listed in Table 1. The results are shown
in Figure 4. As can be seen from Figure 4
the slope of the DCR counts vs. input power does not have a slope of
one as would be expected for a constant gain. This indicates that the
gain is indeed non-linear even for changes of a few Kelvin at the
system temperature of even our best receivers ( K). This also
indicates that the non-linearity does not arise in the receivers since
it is seen from both the X-band and the C-band receivers.
Another test that we did was to observe one source continually while
we varied the attenuation in the IF Rack. This was done to see if the
DCR responded linearly with the change in power coming through the IF
Rack attenuators. We tracked 3C 274 (Virgo A) and balanced the IF Rack
to 1 Volt in each of the RF power samplers for Optical Driver paths 2, 4, 6
and 8. We then added 3 dB to all IF Rack
attenuators (bringing each RF power sampler down to about 0.5 Volts).
We then fired the hi cal ( K) and took data with the DCR being
input from the IF Rack at several different attenuator settings which
are listed in Table 2.
![]() |
The results are plotted in Figure 5. We have plotted the
gain ratio from Equation 5 versus the relative change in
input power (i.e.
). As can be seen in
Figure 5 the gain ratios are not one - indicating
that the DCR response is non-linear with the change in attenuation setting.
This suggests that the observed non-linearity in the system comes after the IF
Rack attenuators.
We can also see from Figure 5 that we have sent the DCR into strong gain compression for the last sets of values. Further investigations are warranted to determine exactly when the DCR goes into gain suppression. This is critical when observing near very bright continuum sources.