| Site 1Hz to: | delay |
|---|---|
| GUPPI | 121 ns |
| GASP | 139 ns |
| Spigot | 107.4 ns |
| SpectralProc | 96.3 ns |
| BCPM (now defunct) | 191.2 ns |
| CGSR2 (now defunct) | 128 ns |
The delay to GASP is 139nS; to CGSR2 128nS. The same procedure was used as described in the web page, except the Racal/Dana counter in the Timing Rack was used instead of the Fluke 6681 Timer/Counter. That should be fine, but it has 1ns resolution. I measured to the end of the NRAO 1pps cables; there are of course delays from these points down into the bowels of the backends.
SiteTime
As shown in the diagram, the master clock generates a 1pps signal, locked to a reference signal from the maser frequency standard. This 1pps, referred to as "SiteTime", is always kept within a few microseconds of UTC by comparison with time signals from a GPS receiver.
Logging delays
The SiteTime 1pps is compared with the GPS signal
and with other 1pps signals by selecting pairs of signals with the
switch and measuring the relative delay with the Racal/Dana counter,
which reads to 1 ns resolution. The switch and counter are computer
controlled, and automatically record the time delays between
SiteTime, GPS, and the other signals. These delays are recorded
about once every two minutes and stored in log files.
These log files in various formats may be found
in the
timing archives.
On 1/31/2001, delays of the 1PPS signal at various points in the Equipment Room were measured. A Fluke 6681 Timer/Counter was used to measure the delays. Two RG223 cables about 25 feet long, and matched to 0.3 nS were used to connect to the test points. The Fluke counter A input was connected to the cable normally connected to the A input of the Racal/Dana counter in the 1PPS Measurement system located in the Timing Center rack. The 1PPS Mux was locked to the X0 input, the Site Time 1PPS from the Trak clock. The Fluke counter B input was then connected to various points in the 1pps distribution system in the Equipment Room:
|