Karen O'Neil
Set-up
Summary
Observations
| Scan | Source | ON time | Comments |
|---|---|---|---|
| 80,81 | UGC 9179 | 300s | RA offset=-87' |
| 82,83 | UGC 9179 | 300s | RA offset=-87' |
| 84,85 | UGC 9179 | 300s | RA offset=-87' |
| 86,87 | 3C295 | 300s | RA offset=-87' |
| 88,89 | 3C295 | 300s | RA offset=-87' |
Analysis
I. Calibrator - 3C295
Gain inferred from this calibrator was 1.76/1.79 & 1.78/1.80 for pol. X/Y and banks A & B, respectively. Identical gains were found for both On+Off observations. RFI was seen in the calibrator scans every 5MHz. The reduced spectrum are shown below.
II. Galaxy - UGC 9179
Three 600s positioned-switched observations were made of U9179. All of the on+off pairs showed extremely flat baselines, as shown below:
The measured flux, peak, and velocity widths of UGC 9179 for all of the On+Off pairs fall within 10% of the value published by Rick Fisher for all banks and all polarizations (Fluxobs = 113-116 Jy km/s; FluxRick=116 Jy km/s; vobs = 307 km/s; vRick=304 km/s). The average of all 3 scans is below:
III. RMS
The rms noise for polarizations X and Y ranged from 0.011 - 0.012K, with no discernable difference between the banks or polarizations. The RMS of the three averaged scans, with all IFS averaged, was 7-9 mK for both polarizations. Assuming a Tsys of 17.5K for the two polarizations (taken from the average values) gives a theoretical value of 13 mK for one 600s scan and 8mK for the average of the four scans. The match is quite good.